+1 713 784 0000  |   KAM RSS Feed Kam on Facebook  
simple precision
Request More Information
 

KF Karl Fischer Moisture Analyzer


KF Karl Fischer Moisture AnalyzerEasy to use and fully automatic, Karl Fischer Moisture Analyzers totally eliminate the troublesome procedures involved with conventional water content determinations, in particular the determination of oil/water content in crude oils. Because they incorporate a special electrolysis current-control system, moisture analyzers ensure the fast and accurate measurement of even trace water content. Moisture analyzers are available in lab (LKF) and portable models (PKF), both of which utilize the coulometric principle applied to Karl Fischer titration.

KAM moisture analyzers rapidly and accurately determine the water content in crude oil samples for all custody transfer operations: production, pipeline, marine, or truck. They can also be used to analyze refined products, transformer oils, jet fuels, chemicals, gas, and most other liquids.


HOW IT WORKS

  • Kam Moisture Analyzers incorporate the coulometric principle applied to Karl Fischer titration. Karl Fischer titrimetry is an accurate moisture measurement method utilizing the quantitative reaction of water with iodine.

KEY ADVANTAGES

  • Fully automatic operation
  • Automatic electrolysis control and blank current control system
  • No reagent calibration and no burettes required
  • 5-digit digital display
  • Bar-graph meters monitor progress of titration
  • Titration cell assembly
  • Portable model has 10 hours continuous operation on rechargeable battery

APPLICATIONS

  • Custody transfer
  • Water concentration in refined fuels, including aviation
  • Water concentration in machine oils
  • Water concentration in hydrofluoric acid
  • Water concentration in gas

KF Karl Fischer Moisture Analyzer - Specifications


PER API MPMS 10.9, ASTM D4928, and EI 386

Method: Coulometric Karl Fischer titration
Detection: Polarization detection
Control: Automatic electrolysis current control
Display: 5-digit display, μg H2O or 0.1, 0.2, 0.5, 1.0 ml volume % H2O
Sample size: 0.1, 0.2, 0.5, 1.0 ml for direct reading of volume % H2O, or less than 2 grams (or ml)
Range: 10μg - 100,000μg H2O (i.e. for a 100μg sample, the direct display is 0.0010% to 10% H2O)
Sensitivity: 1 μg H2O
Precision: ±5μg for 10μg - 1000mg, 0.5% (C.V.) for over 1000μg (meets or exceeds API MPMS 10.9, ASTM D4928, and EI 386)
Titration speed: 1000lg H2O/min. (max. at high H2O concentrations)
Power requirements: Operates on either AC or DC. AC - 110/120, 220/240 V, 50/60Hz  DC - Rechargeable battery for portable model (12-14V)
Dimensions: Approx. 15” x 10”x 9” (381mm x 254mm x 229mm)
Weight: Lab model - approx. 10 lbs. (4.5 kg) Portable model - approx. 20 lbs. (9 kg)

KF Karl Fischer Moisture Analyzer - Models




KF Karl Fischer Lab





KF Karl Fischer Portable





 

Website Design by FirstLink Technology, Inc.